A dissolution test is a means of identifying and proving the availability of active pharmaceutical ingredient (API) in their delivered form.
The DFC offline semi-automated dissolution testing systems include a dissolution bath, a pump and a fraction collector which collects the sampled dissolution media.
Automated dissolution testing in this case means sampling, measuring, and result calculation are automated. The user adds the samples when prompted to do so by the system and takes care of filling and cleaning of the vessels.
Media preparation and degassing are vital parts of a tablet dissolution test as deaerated media are critical to the reproducibility of dissolution results.
UV/VIS Spectrometers for Tablet Dissolution Applications
Media Preparation and Degassing System
Media degassing or “de-aeration”, pre-heating to a temperature of 37°C and absolutely precise gravimetric dispensing according to the recommendations and demanded for dissolution testing from the US FDA authorities are standard for the Pharma Test PT-DDS4 media preparation and degasing system. De-aeration is generally recommended to prevent the formation of bubbles on the dosage form surface, which may cause a variation in active dissolution rates, by virtue of surface obscuration.
Parameter | Specification |
Dosing Range | 250 – 5,000g – gravimetric method
250 – 5,000ml – volumetric method |
Dosing Accuracy | ±0.5% of selected volume – gravimetric method
±0.8% of selected volume – volumetric method |
Tank Volume | Approx. 24 liter |
Filling Volume | Not less than min. 10 liter |
Dispensing Time for 1 Liter | Approx. 45 – 60 seconds |
De-aeration and Heating Time | 24 liters within approx. 30 – 45 minutes |
Display | LED backlit LCD, 320×240 pixels, 9″ diagonal, approx. 90x120mm |
Data Entry | Alpha numerical keypad |
Method Memory | Store up to 16 methods |
Printer | Thermo printer, 24 characters |
Interface | RS232 port |
Vacuum Level | 0.4 bar absolute |
Temperature Range | 30.0 – 50.0°C |
Temperature Accuracy | ±0.5°C |
Heater Capacity | 900W |
Calibration | Built-in calibration procedures for digital load cell (weight), digital flow meter (volume) temperature control |
Height of Dispenser | 130cm from ground floor |
Space Requirements | 550 x 700 x 900 mm (Length x Width x Height) |
Packaging Dimensions | Approx. 1070 x 970 x 780cm (Length x Width x Height) |
Net / Gross weight | Approx. 75 / 110 kg |
Certification | All components certified to USP / EP requirements |
CE / EMC Certification | All CE / EMC Certification provided |
Validation | All IQ & OQ documents included |