edge electrodes are highly advanced, featuring a built-in microchip that stores calibration information that is automatically retrieved by edge once the electrode is plugged in. These digital electrodes also feature an easy to plug in 3.5 mm connector so you never have to worry about the right angle or aligning pins.
When used with Hanna’s pH electrodes equipped with a matching pin, edge constantly checks the impedance of the pH measuring electrode to notify you in real time in the event of glass breakage. During calibration, Sensor Check checks the state of the junction. The reference junction is also evaluated and reported on the display.
Plugging an electrode in has never been simpler; no alignments, no broken pins, just connect the 3.5 mm plug and get started.
edge® is incredibly versatile. It can be used in a variety of ways and in a variety of environments without taking up the space of a traditional benchtop meter.
Using the wall mount cradle (included), edge can be placed on a wall, leaving zero footprint on the benchtop space. The cradle has a built-in connector to power edge and charge its batteries. edge’s zero footprint is designed to save you valuable benchtop space.
edge is equipped with a benchtop cradle with an adjustable swivel electrode holder to charge and hold the edge securely in place at the optimum viewing angle.
edge® is incredibly thin and weighs only 8.8 oz., has an 8 hour battery life, and dual USB inputs.
edge includes one standard USB for exporting data with a flash drive and a second micro USB port for you to connect to a computer for file export and for charging your edge when the cradle is not available.
edge allows you to store up to 1000 log records of data. Data sets include readings, GLP data, date and time.
edge can be used in Extended or Basic Operating Modes. Extended Mode enables all features available while the Basic Mode reduces the amount of features to a very simple operation. For example, in extended mode edge offers 5 point calibration with 2 custom buffers while in basic mode calibration can only be performed to 3 points with 5 preprogrammed buffers.
HI2020 pH kit includes: HI11310 glass body, refillable pH electrode with temperature sensor, pH 4 buffer solution sachets, pH 7 buffer solution sachets, pH 10 buffer solution sachets, electrode cleaning solution sachets,USB cable, benchtop docking station with electrode holder, wall mount cradle, 5 VDC power adapter, instruction manual, and quality certificate.
SKU | HI2020-01 |
Name | ? Multiparameter pH Meter – HI2020 |
pH Range | -2.000 to 16.000 pH, -2.00 to 16.00 pH, ?1000 mV |
pH Resolution | 0.001 pH, 0.01 pH, 0.1 mV |
pH Accuracy | ?0.002 pH, ?0.01 pH, ?0.2 mV |
pH Calibration | 5 points (Standard mode) 1.68, 4.01 (3.00?), 6.86, 7.01, 9.18, 10.01, 12.45, and two custom buffers; 3 points (Basic mode) 4.01; 6.86; 7.01; 9.18; 10.01 |
pH Temperature Compensation | ATC: -5.0 to 100.0?C; 23.0 to 212.0?F* |
pH CAL Check (electrode diagnostics) | Glass and reference junction diagnostics ( HI 11311 & HI 12301 only), out of calibration range , probe condition, response time |
EC Range | 0.00 to 29.99 µS/cm, 30.0 to 299.9 µS/cm, 300 to 2999 µS/cm, 3.00 to 29.99 mS/cm, 30.0 to 200.0 mS/cm, up to 500.0 mS/cm (absolute EC)** |
EC Resolution | 0.01 µS/cm, 0.1 µS/cm, 1 µS/cm, 0.01 mS/cm, 0.1 mS/cm |
EC Accuracy | ?1% of reading (?0.05 µS/cm or 1 digit, whichever is greater) |
EC Calibration | 1 point offset calibration (0.00 µS/cm in air), 1 point slope calibration in EC standard 84 µS/cm, 1413 µS/cm, 5.00 mS/cm, 12.88 mS/cm, 80.0 mS/cm and 111.8 mS/cm |
TDS Range | 0.00 to 14.99 ppm (mg/L), 15.0 to 149.9 ppm (mg/L), 150 to 1499 ppm (mg/L), 1.50 to 14.99 g/L, 15.0 to 100.0 g/L, up to 400.0 g/L (absolute TDS)**, with 0.80 conversion factor |
TDS Resolution | 0.01 ppm, 0.1 ppm, 1 ppm, 0.01 g/L, 0.1 g/L |
TDS Accuracy | ?1% of reading (?0.03 ppm or 1 digit, whichever is greater) |
TDS Calibration | through EC calibration |
EC/TDS Temperature Compensation | ATC (0.0 to 100.0?C; 32.0 to 212.0 ?F), NoTC |
EC to TDS Conversion Factor | 0.40 to 0.80 |
Salinity (% NaCl) Range | 0.0 to 400.0% NaCl*** |
Salinity (% NaCl) Resolution | 0.1% NaCl |
Salinity (% NaCl) Accuracy | ?1% of reading (@25?C/77?F) |
Salinity (% NaCl) Calibration | 1 point with HI 7037L 100% NaCl sea water standard |
Salinity (PSU) Range | 2.00 to 42.00 PSU*** |
Salinity (PSU) Resolution | 0.01 PSU |
Salinity (PSU) Accuracy | ?1% of reading (@25?C/77?F) |
Salinity (PSU) Calibration | through EC calibration |
Salinity Range | 0.0 to 80.0 g/L*** |
Salinity Resolution (ppt) | 0.01 g/L |
Salinity Accuracy (ppt) | ?1% of reading (@25?C/77?F) |
Salinity Calibration (ppt) | through EC calibration |
Dissolved Oxygen Range | 0.00 to 45.00 ppm (mg/L), 0.0 up to 300.0% saturation |
Dissolved Oxygen Resolution | 0.01 ppm, 0.1% saturation |
Dissolved Oxygen Accuracy | ?1 digit, ?1.5% of reading |
Dissolved Oxygen Calibration | one or two points at 0% (HI 7040 solution) and 100% (in air) |
Dissolved Oxygen Temperature Compensation | ATC: 0 to 50 ?C; 32.0 to 122.0 ?F |
Salinity Compensation | 0 to 40 g/L (with 1 g/L resolution) |
Altitude Compensation | -500 to 4000 m (with 100 m resolution) |
Temperature Range | -20.0 to 120.0 ?C, -4.0 to 248.0 ?F |
Temperature Resolution | 0.1 ?C, 0.1 ?F |
Temperature Accuracy | ?0.5 ?C, ?0.9 ?F |
pH Electrode | HI11310 pH electrode (included) |
Logging Type | Manual log-on-demand (Max. 200 logs); Manual log-on-stability (Max. 200 logs); Interval logging *** (Max. 600 samples; 100 lots) |
Logging Memory | up to 1000*** records |
Connectivity | 1 micro USB port for charging and PC connectivity, 1 USB port for storage |
Battery Type/Life | Built-in rechargeable battery with up to 8 hours of continuous use |
Power Supply | 5 VDC adapter (included) |
Environment | 0 to 50?C (32 to 122?F), RH max 95% non-condensing |
Dimensions | 202 x 140 x 12.7mm (8” x 5.5” x 0.5”) |
Weight | 250 g (8.82 oz) |
The HI5522 is an advanced research grade benchtop pH/ORP/ISE and EC/TDS/Salinity/Resistivity meter that is completely customizable with a large color LCD, capacitive touch keys, and USB port for computer connectivity.
The HI5522 is a two-channel meter that allows for simultaneous measure of pH, ORP, or ISE on one channel and EC, TDS, Salinity, or Resistivity on the other. Channel 1 has a BNC connection for use with the expansive line of pH, ORP, and ISE electrodes that Hanna Instruments offers. The meter is supplied with the HI1131B glass body, double junction, combination pH electrode that operates over a wide temperature range from 0 to 100 ℃. All readings are automatically compensated for temperature variations with the separate HI7662-T temperature probe or from the built in temperature sensor of the conductivity probe on Channel 2. The HI5522 is supplied with the HI76312 four-ring conductivity probe that operates over a wide range from 0.000 μS/cm to 1000.0 mS/cm*. The meter can be set to auto-ranging in which the meter chooses the appropriate conductivity range from seven ranges or fixed range in which the meter will only display reading in μS/cm or mS/cm. All readings are automatically compensated for temperature variations with a built in temperature sensor. The temperature correction coefficient is adjustable from 0.00 to 10.00 %/℃.
As a pH meter the HI5522 can be calibrated up to five points with a choice of eight pre-programmed buffers or five custom buffers. The HI5522 features Hanna’s exclusive CAL Check™ to alert the user of potential problems during the pH calibration process. Indicators displayed during calibration include “Electrode Dirty/Broken” and “Buffer Contaminated.” The overall probe condition based on the offset and slope characteristic of the electrode is displayed as a percentage after calibration is complete.
In ISE mode the HI5522 can be calibrated up to five points with a choice of five fixed standards or five user defined in any concentration unit. The calibration data including date, time, standards used and slope can be viewed at any time along with the current measurement by selecting the Good Laboratory Practice (GLP) display option.
As an EC/TDS/Salinity/Resistivity meter the HI5522 can be calibrated up to four points with a choice of six pre-programmed conductivity standards or user defined custom standards. Resistivity, TDS, Practical Salinity (PSU) and Natural Seawater Scale are calibrated through conductivity. The % NaCl is calibrated to single point with the HI7037 salinity standard. The calibration data including date, time, and standards used, offset and cell factor can be accessed at any time along with the current measurement by selecting the Good Laboratory Practice (GLP) display option.
For the measurement of high purity water used in pharmaceutical manufacturing, the HI5522 is programmed with the three stages of the USP <645> method. Once a stage is met a report is generated and can be saved. Up to 200 reports can be stored and transferred to a Windows® compatible computer using a USB cable and software (sold separately).
Three selectable logging modes are available: automatic, manual and AutoHold logging. Up to 100, 000 data points can be recorded in 100 lots with 50, 000 records max/lot on each channel and exported to a computer for data review and storage.
Highly Customizable User Interface – The user interface of the HI5522 allows the user to show measurements in various modes: basic measurement with or without GLP information, real-time graphing, and logging data. Calibration stability criteria can be adjusted from fast, moderate, and accurate. Programmable alarm limits can be set to inside or outside allowable limits.
Color Graphic LCD – The HI5522 features a color graphic LCD with on-screen help, graphic, and custom color configurations. The display allows for real-time graphing and the use of virtual keys provide for an intuitive user interface.
Capacitive Touch – The HI5522 features sensitive capacitive touch buttons for accurate keystrokes when navigating menus and screens. There are four dedicated keys that are used for routine operations including calibration and switching measurement modes and four virtual keys that change based upon use. The capacitive touch technology ensures the buttons never get clogged with sample residue.
Four Ring Conductivity Probe – All readings are performed with the HI76312 four-ring conductivity probe that has a built in temperature sensor for automatic temperature correction. The four rings are made with platinum and the body of the electrode is made of Polyetherimide (PEI) plastic that is resistant to many harsh chemicals. The four-ring design allows for this probe to be used over a wide range of measurements.
Choice of Calibration – Automatic buffer recognition, semiautomatic, and direct manual entry pH calibration options are available for calibrating up to five points, from a selection of eight standard buffers and up to five custom buffers. For the conductivity channel the calibration can be set to automatic standard recognition or user entry along with a choice of single or multi-point. Calibration can be performed up to four points when multi-point is selected.
CAL Check™– CAL Check™ alerts users to potential problems during the calibration of the pH electrode. Indicators include “Electrode Dirty/Broken, ” “Buffer Contaminated, ” electrode response timeand the overall probe condition as a percentage that is based on the offset and slope characteristics.
GLP Data – HI5522 includes a GLP Feature thatallows users to view calibration data and calibration expiration information at the touch of a key. Calibration data include date, time, standards used for calibration.
ISE Measurement with Choice of Concentration Units – The HI5522 allows for calibration and readings in choice of concentration units. The choices of concentration units include ppt, g/L, mg/mL, ppm, mg/L, μg/L, ppb, μg/L, mg/mL, M, mol/L, mmol/L, %w/v and a user-defined unit.
ISE Measurement with Incremental Methods – The known addition, known subtraction, analyte addition, and analyte subtraction incremental methods are pre-programmed into the HI5522. Simply follow the on screen guided procedure and the meter will perform the calculation automatically allowing for a higher level of accuracy to be obtained as compared to a direct ISE measurement.
Data Logging – Three selectable logging modes are available on the HI5522: automatic, manual, and AutoHold logging. Automatic and manual logs up to 100 lots with 50, 000 records max/lot, with up to 100, 000 total data points. Automatic logging features the option to save data according to sampling period and interval.
Data Transfer – Data can be transferred to a PC with USB cable and HI92000 software (both sold separately).
Contextual Help – Contextual help is always available through a dedicated “HELP” key. Clear tutorial messages and directions are available on-screen to quickly and easily guide users through setup and calibration. The help information displayed is relative to the setting/option being viewed.
CAL Check alert to potential problems during calibration
Comprehensive GLP Data
Real time logging
HI5522 is supplied with HI1131 pH electrode, HI7662-T temperature probe, HI76312 conductivity probe, HI76404W electrode holder, HI70004 pH 4.01 buffer solution sachet, HI70007 pH 7.01 buffer solution sachet, HI700601 electrode cleaning solution sachet (2), HI7082 3.5M KCl electrolyte solution (30 mL), 12 VDC power adapter and instructions.
SKU | HI5522-02 |
---|---|
Product Name | Laboratory Research Grade Benchtop pH/mV/ISE and EC/TDS/Salinity/Resistivity Meter – HI5522 |
Quote Required | Yes |
pH Range | -2.0 to 20.0 pH / -2.00 to 20.00 pH / -2.000 to 20.000 pH |
pH Resolution | 0.1 pH / 0.01 pH / 0.001 pH |
pH Accuracy | ±0.1 pH / ±0.01 pH / ±0.002 pH ± 1LSD |
pH Calibration | Up to five-point calibration, eight standard buffers available (pH 1.68, 3.00, 4.01, 6.86, 7.01,9.18, 10.01, 12.45), and five custom buffers |
pH Temperature Compensation | automatic or manual from -20.0 to 120.0 °C |
mV Range | ±2000.0 mV |
mV Resolution | 0.1 mV |
mV Accuracy | ±0.2 mV ± 1LSD |
mV Relative Offset | ±2000.0 mV |
ISE Range | e.g. 10-7 to 10M/ 0.005 to 105 ppm/ 5·10–7 to 5·107 conc |
ISE Resolution | 1 conc./ 0.1 conc. / 0.01 conc. / 0.001 conc |
ISE Accuracy | ±0.5% (monovalent ions); ±1% (divalent ions) |
ISE Calibration | Up to five-point calibration, seven fixed standard solutions available for each measurement unit, and five custom solutions |
EC Range | 0.000 to 9.999 μS/cm, 10.00 to 99.99 μS/cm, 100.0 to 999.9 μS/cm, 1.000 to 9.999 mS/cm, 10.00 to 99.99 mS/cm, 100.0 to 1000.0 mS/cm actual EC* |
EC Resolution | 0.001 μS/cm, 0.01 μS/cm, 0.1 μS/cm, 1 μS/cm, 0.001 mS/cm, 0.01 mS/cm, 0.1 mS/cm |
EC Accuracy | ±1% of reading (±0.01 μS/cm) |
EC Calibration | automatic standard recognition (0.000 μS/cm, 84.00 μS/cm, 1.413 mS/cm, 5.000 mS/cm, 12.88 mS/cm, 80.00 mS/cm, 111.8 mS/cm) or user standard; single point or multi-point calibration |
TDS Range | 0.000 to 9.999 ppm, 10.00 to 99.99 ppm, 100.0 to 999.9 ppm, 1.000 to 9.999 ppt, 10.00 to 99.99 ppt, 100.0 to 400.0 ppt actual TDS* (with 1.00 factor) |
TDS Resolution | 0.001 ppm, 0.01 ppm, 0.1 ppm, 1 ppm, 0.001 ppt, 0.01 ppt, 0.1 ppt |
TDS Accuracy | ±1% of reading (±0.01 ppm) |
EC/TDS Temperature Compensation | Disabled / Linear / Non linear (natural water) |
Temperature Correction Coefficient | 0.00 to 10.00 %/°C |
Reference Temperature | 5.0 °C to 30.0 °C |
Cell Constant | 0.0500 to 200.00 |
Cell Type | 4 cells |
Salinity (% NaCl) Range | 0.0 to 400.0% |
Salinity (% NaCl) Resolution | 0.01% |
Salinity (% NaCl) Accuracy | ±1% of reading |
Salinity (PSU) Range | 0.00 to 42.00 psu |
Salinity (PSU) Resolution | 0.01 |
Salinity (PSU) Accuracy | ±1% of reading |
Salinity (PSU) Calibration | one-point (with HI7037 standard) |
Salinity Range | 0.00 to 80.00 ppt |
Salinity Resolution (ppt) | 0.01 |
Salinity Accuracy (ppt) | ±1% of reading |
Resistivity Range | 1.0 to 99.9 Ω•cm; 100 to 999 Ω•cm; 1.00 to 9.99 KΩ•cm; 10.0 to 99.9 KΩ•cm; 100 to 999 KΩ•cm; 1.00 to 9.99 MΩ•cm; 10.0 to 100.0 MΩ•cm |
Resistivity Resolution | 0.1 Ω•cm; 1 Ω•cm; 0.01 KΩ•cm; 0.1 KΩ•cm; 1 KΩ•cm; 0.01 MΩ•cm; 0.1 MΩ•cm* |
Resistivity Accuracy | ±2% of reading (±1 Ω•cm) |
Temperature Range | -20.0 to 120.0 ºC**; -4.0 to 248.0 ºF**; 253.15 to 393.15 K** |
Temperature Resolution | 0.1 °C, 0.1 °F, 0.1 K |
Temperature Accuracy | ±0.2 °C; ±0.4 °F; ±0.2 K (without probe) |
Input Channels | 2 (pH/mV/ISE; Conductivity/Resistivity/TDS/ Salinity) |
pH Electrode | HI1131B glass body pH electrode with BNC connector and 1 m (3.3’) cable (included) |
EC/TDS Probe | HI76312 platinum, four-ring EC/TDS probe with and 1 m (3.3’) cable (included) |
Temperature Probe | HI7662-T stainless steel temperature probe with 1 m (3.3’) cable (included) |
Logging Type | Automatic, Manual, AutoHold |
Logging Memory | Up to 100 lots, 50,000 records max/lot; maximum 100,000 data points / channel |
Logging Interval | 14 selectable between 1 second and 180 minutes |
Connectivity | Opto-isolated USB |
GLP | yes (GLP Channel 1 |
Calibration Timeout Reminder | yes |
USP Compliant | USP <645> Application |
Profiles | Up to 10 (5 for each) |
Display | Color Graphic LCD 240 x 320 pixels, 8 keys capacitive touch keyboard |
Power Supply | 12 VDC adapter (included) |
Environment | 0 to 50°C (32 to 122°F; 273 to 323 K), RH max 95% non-condensing |
Dimensions | 160 x 231 x 94 mm (6.3 x 9.1 x 3.7”) |
Weight | 1.2 Kg (2.6 lbs) |
Ordering Information | HI5522 is supplied with HI1131 pH electrode, HI7662-T temperature probe, HI76312 conductivity probe, HI76404W electrode holder, HI70004 pH 4.01 buffer solution sachet, HI70007 pH 7.01 buffer solution sachet, HI700601 electrode cleaning solution sachet (2), HI7082 3.5M KCl electrolyte solution (30 mL), 12 VDC power adapter and instructions. |
Notes | *Uncompensated conductivity (or TDS) is the conductivity (or TDS) value without temperature compensation. |
The HI5521 is an advanced research grade benchtop pH/ORP and EC/TDS/Salinity/Resistivity meter that is completely customizable with a large color LCD, capacitive touch keys, and USB port for computer connectivity.
The HI5521 is a two-channel meter that allows for simultaneous measure of pH or ORP on one channel and EC or related parameters on the other. Channel 1 has a BNC connection for use with the expansive line of pH and ORP electrodes that Hanna Instruments offers. The meter is supplied with the HI1131B glass body, double junction, combination pH electrode that operates over a wide temperature range from 0 to 100 ℃. All readings are automatically compensated for temperature variations with the separate HI7662-T temperature probe or from the built in temperature sensor of the conductivity probe on Channel 2. The HI5521 is supplied with the HI76312 four-ring conductivity probe that operates over a wide range from 0.000 μS/cm to 1000.0 mS/cm*. The meter can be set to auto-ranging in which the meter chooses the appropriate conductivity range from seven ranges or fixed range in which the meter will only display reading in μS/cm or mS/cm. All readings are automatically compensated for temperature variations with a built in temperature sensor. The temperature correction coefficient is adjustable from 0.00 to 10.00 %/℃.
As a pH meter the HI5521 can be calibrated up to five points with a choice of eight pre-programmed buffers or five custom buffers. The HI5521 features Hanna’s exclusive CAL Check™ to alert the user of potential problems during the pH calibration process. Indicators displayed during calibration include “Electrode Dirty/Broken” and “Buffer Contaminated.” The overall probe condition based on the offset and slope characteristic of the electrode is displayed as a percentage after calibration is complete. The calibration data including date, time, buffers used, offset and slope can be accessed at any time along with the current measurement by selecting the Good Laboratory Practice (GLP) display option.
As an EC/TDS/Salinity/Resistivity meter the HI5521 can be calibrated up to four points with a choice of six pre-programmed conductivity standards or user defined custom standards. Resistivity, TDS, Practical Salinity (PSU) and Natural Seawater Scale are calibrated through conductivity. The % NaCl is calibrated to single point with the HI7037 salinity standard. The calibration data including date, time, and standards used, offset and cell factor can be accessed at any time along with the current measurement by selecting the Good Laboratory Practice (GLP) display option.
For the measurement of high purity water used in pharmaceutical manufacturing, the HI5521 is programmed with the three stages of the USP <645> method. Once a stage is met a report is generated and can be saved. Up to 200 reports can be stored and with the USB port be transferred to a Windows® compatible computer.
Three selectable logging modes are available: automatic, manual and AutoHold logging. Up to 100, 000 data points can be recorded in 100 lots with 50, 000 records max/lot on each channel and exported to a computer for data review and storage.
Highly Customizable User Interface – The user interface of the HI5521 allows the user to show measurements in various modes: basic measurement with or without GLP information, real-time graphing, and logging data. Calibration stability criteria can be adjusted from fast, moderate, and accurate. Programmable alarm limits can be set to inside or outside allowable limits.
Color Graphic LCD – The HI5521 features a color graphic LCD with on-screen help, graphic, and custom color configurations. The display allows for real-time graphing and the use of virtual keys provide for an intuitive user interface.
Capacitive Touch – The HI5521 features sensitive capacitive touch buttons for accurate keystrokes when navigating menus and screens. There are four dedicated keys that are used for routine operations including calibration and switching measurement modes and four virtual keys that change based upon use. The capacitive touch technology ensures the buttons never get clogged with sample residue.
Four Ring Conductivity Probe – All readings are performed with the HI76312 four-ring conductivity probe that has a built in temperature sensor for automatic temperature correction. The four rings are made with platinum and the body of the electrode is made of Polyetherimide (PEI) plastic that is resistant to many harsh chemicals. The four-ring design allows for this probe to be used over a wide range of measurements.
Choice of Calibration – Automatic buffer recognition, semiautomatic, and direct manual entry pH calibration options are available for calibrating up to five points, from a selection of eight standard buffers and up to five custom buffers. For the conductivity channel the calibration can be set to automatic standard recognition or user entry along with a choice of single or multi-point. Calibration can be performed up to four points when multi-point is selected.
CAL Check™ – CAL Check™ alerts users to potential problems during the calibration of the pH electrode. Indicators include “Electrode Dirty/Broken, ” “Buffer Contaminated, ” electrode response timeand the overall probe condition as a percentage that is based on the offset and slope characteristics.
GLP Data – HI5521 includes a GLP Feature thatallows users to view calibration data and calibration expiration information at the touch of a key. Calibration data include date, time, standards used for calibration.
Data Logging – Three selectable logging modes are available on the HI5521: automatic, manual, and AutoHold logging. Automatic and manual logs up to 100 lots with 50, 000 records max/lot, with up to 100, 000 total data points. Automatic logging features the option to save data according to sampling period and interval.
Data Transfer – Data can be transferred to a PC with USB cable and HI92000 software (both sold separately).
Contextual Help – Contextual help is always available through a dedicated “HELP” key. Clear tutorial messages and directions are available on-screen to quickly and easily guide users through setup and calibration. The help information displayed is relative to the setting/option being viewed.
CAL Check alert to potential problems during calibration
Comprehensive GLP Data
Real time logging
HI5521 is supplied with HI1131 pH electrode, HI7662-T temperature probe, HI76312 conductivity probe, HI76404W electrode holder, HI70004 pH 4.01 buffer solution sachet, HI70007 pH 7.01 buffer solution sachet, HI700601 electrode cleaning solution sachet (2), HI7082 3.5M KCl electrolyte solution (30 mL), 12 VDC power adapter and instructions.
SKU | HI5521-02 |
---|---|
Product Name | Laboratory Research Grade Benchtop pH/mV and EC/TDS/Salinity/Resistivity Meter – HI5521 |
Quote Required | Yes |
pH Range | -2.0 to 20.0 pH / -2.00 to 20.00 pH / -2.000 to 20.000 pH |
pH Resolution | 0.1 pH / 0.01 pH / 0.001 pH |
pH Accuracy | ±0.1 pH / ±0.01 pH / ±0.002 pH ± 1LSD |
pH Calibration | Up to five-point calibration, eight standard buffers available (pH 1.68, 3.00, 4.01, 6.86, 7.01,9.18, 10.01, 12.45), and five custom buffers |
pH Temperature Compensation | automatic or manual from -20.0 to 120.0 °C |
mV Range | ±2000.0 mV |
mV Resolution | 0.1 mV |
mV Accuracy | ±0.2 mV ± 1LSD |
mV Relative Offset | ±2000.0 mV |
EC Range | 0.000 to 9.999 μS/cm, 10.00 to 99.99 μS/cm, 100.0 to 999.9 μS/cm, 1.000 to 9.999 mS/cm, 10.00 to 99.99 mS/cm, 100.0 to 1000.0 mS/cm actual EC* |
EC Resolution | 0.001 μS/cm, 0.01 μS/cm, 0.1 μS/cm, 1 μS/cm, 0.001 mS/cm, 0.01 mS/cm, 0.1 mS/cm |
EC Accuracy | ±1% of reading (±0.01 μS/cm) |
EC Calibration | automatic standard recognition (0.000 μS/cm, 84.00 μS/cm, 1.413 mS/cm, 5.000 mS/cm, 12.88 mS/cm, 80.00 mS/cm, 111.8 mS/cm) or user standard; single point or multi-point calibration |
TDS Range | 0.000 to 9.999 ppm, 10.00 to 99.99 ppm, 100.0 to 999.9 ppm, 1.000 to 9.999 ppt, 10.00 to 99.99 ppt, 100.0 to 400.0 ppt actual TDS* (with 1.00 factor) |
TDS Resolution | 0.001 ppm, 0.01 ppm, 0.1 ppm, 1 ppm, 0.001 ppt, 0.01 ppt, 0.1 ppt |
TDS Accuracy | ±1% of reading (±0.01 ppm) |
EC/TDS Temperature Compensation | disabled, linear and non-linear (natural water) |
Temperature Correction Coefficient | 0.00 to 10.00 %/°C |
Reference Temperature | 5.0 to 30.0 °C |
Cell Constant | 0.0500 to 200.00 |
Cell Type | 4 cells |
Salinity (% NaCl) Range | 0.0 to 400.0 % |
Salinity (% NaCl) Resolution | 0.1 % |
Salinity (% NaCl) Accuracy | ±1% of reading |
Salinity (PSU) Range | 0.00 to 42.00 psu |
Salinity (PSU) Resolution | 0.01 |
Salinity (PSU) Accuracy | ±1% of reading |
Salinity (PSU) Calibration | 1 point (with HI7037 standard) |
Salinity Range | 0.00 to 80.00 ppt |
Salinity Resolution (ppt) | 0.01 |
Salinity Accuracy (ppt) | ±1% of reading |
Resistivity Range | 1.0 to 99.9 Ω•cm; 100 to 999 Ω•cm; 1.00 to 9.99 KΩ•cm; 10.0 to 99.9 KΩ•cm; 100 to 999 KΩ•cm; 1.00 to 9.99 MΩ•cm; 10.0 to 100.0 MΩ•cm |
Resistivity Resolution | 0.1 Ω•cm; 1 Ω•cm; 0.01 KΩ•cm; 0.1 KΩ•cm; 1 KΩ•cm; 0.01 MΩ•cm; 0.1 MΩ•cm* |
Resistivity Accuracy | ±2% of reading (±1 Ω•cm) |
Temperature Range | -20.0 to 120.0 ºC**; -4.0 to 248.0 ºF**; 253.15 to 393.15 K** |
Temperature Resolution | 0.1 °C, 0.1 °F, 0.1 K |
Temperature Accuracy | ±0.2 °C; ±0.4 °F; ±0.2 K (without probe) |
Input Channels | 2 (pH/mV; Conductivity/Resistivity/ TDS/Salinity) |
pH Electrode | HI1131B glass body pH electrode with BNC connector and 1 m (3.3’) cable (included) |
EC/TDS Probe | HI76312 platinum, four-ring EC/TDS probe with and 1 m (3.3’) cable (included) |
Temperature Probe | HI7662-T stainless steel temperature probe with 1 m (3.3’) cable (included) |
Logging Type | Automatic, Manual, AutoHold |
Logging Memory | Up to 100 lots, 50,000 records max/lot; maximum 100,000 data points / channel |
Logging Interval | 14 selectable between 1 second and 180 minutes |
Connectivity | Opto-isolated USB |
GLP | yes (GLP Channel 1 |
Calibration Timeout Reminder | yes |
USP Compliant | USP <645> Application |
Profiles | up to 10, 5 each channel |
Display | Color Graphic LCD 240 x 320 pixels, 8 keys capacitive touch keyboard |
Power Supply | 12 VDC adapter (included) |
Environment | 0 to 50°C (32 to 122°F; 273 to 323 K), RH max 95% non-condensing) |
Dimensions | 160 x 231 x 94 mm (6.3 x 9.1 x 3.7”) |
Weight | 1.2 Kg (2.6 lbs |
Ordering Information | HI5521 is supplied with HI1131 pH electrode, HI7662-T temperature probe, HI76312 conductivity probe, HI76404W electrode holder, HI70004 pH 4.01 buffer solution sachet, HI70007 pH 7.01 buffer solution sachet, HI700601 electrode cleaning solution sachet (2), HI7082 3.5M KCl electrolyte solution (30 mL), 12 VDC power adapter and instructions. |
Notes | *Uncompensated conductivity (or TDS) is the conductivity (or TDS) value without temperature compensation. |