6-position thermoreactor for Chemical Oxygen Demand (COD) with settable temperature (up to 200 °C) and timer (up to 199 minutes).
COD Analysis Thermoreactor ECO 6 ensures a simple and easy programming of time (from 1 to 199 minutes or continuous) and temperature (from ambient to 200 °C).
Through the digital display, it is possible to monitor the test at any time, with the countdown constantly shown.
The unit automatically shuts down when the analysis time expires, whilst safety is of primary importance, as demonstrated by the antisplash bells that protect the user and other samples and the several acoustic and visual signals that report unpleasant situations.
CONSTRUCTION MATERIAL: | Epoxy painted metal structure |
HOLES NUMBER AND DIAMETER: | 6 positions (Ø 42 mm) |
SET TEMPERATURE: | display visualization |
COUNTDOWN: | display visualization |
END OF THE CYCLE: | acoustic signal with automatic switch-off |
POWER: | 700 W |
WEIGHT: | 5.6 Kg (12.3 lb) |
DIMENSIONS UNCRATED (WXHXD): | 198x132x319mm (7.8×5.2×12.6 in) |
DIMENSIONS CRATED (WXHXD): | 400x280x270 mm |
SELECTABLE WORKING TEMPERATURES: | from room temperature to 200 °C, resolution 1 °C |
SELECTABLE WORKING TIMES: | from 0 to199 minutes or continuous operation |
TEMPERATURE STABILITY: | ± 0.5 °C |
TEMPERATURE HOMOGENEITY: | ± 0.5 °C |
TEMPERATURE PRECISION: | ± 1 °C |
REACHING OF SET TEMPERATURE: | visual signal |
COUNTDOWN: | visual signal |
END OF CYCLE: | acoustic and visual signal |
PROBE INTERRUPTION: | acoustic and visual signal |
BROKEN PROBE: | acoustic and visual signal |
EXCEED TEMPERATURE RANGE: | acoustic and visual signal |
8- and 25-position thermoreactors for Chemical Oxygen Demand (COD) with settable temperature (70, 100, 120, 150, 160 °C) and timer (30, 60, 120 minutes or continuous).
The ECO 8 and ECO 25 provide easy and intuitive programming of both time and temperature, the two parameters required to perform analysis.
The user can switch from one value to another, being able to select between five temperatures (70, 100, 120, 150 and 160 °C) and four different timings (30, 60, 120 minutes or continuous).
The thermoreaction in small test tubes minimizes reagent consumption and reduces the equipment required, as only one heating block is needed to operate on multiple samples simultaneously.
Each test features a disposable vial that is inserted into the block, and requires small amounts of reagent and sample, lowering costs.
The unit automatically shuts down when the analysis time expires.
CONSTRUCTION MATERIAL: | Epoxy painted metal structure |
MODELS: | ECO 8 8 pos. (Ø16 mm) + 1 pos. (Ø22 mm) ECO 25 25 pos. (Ø16 mm) |
SET TEMPERATURE: | display visualization |
COUNTDOWN: | display visualization |
END OF THE CYCLE: | acoustic signal with automatic switch-off |
POWER: | ECO 8 140 W ECO 25 400 W |
WEIGHT: | ECO 8 2 Kg (4.4 lb) ECO 25 3.8 Kg (8.4 Ib) |
DIMENSIONS (WXHXD): | ECO 8 135x95x230 mm (5.3×3.7×9.1 in) ECO 25 155x95x275 mm (6.1×3.7×10.8 in) |
SELECTABLE WORKING TEMPERATURES: | 70, 100, 120, 150 and 160 |
SELECTABLE WORKING TIMES: | 30, 60, 120 or continuous |
TEMPERATURE STABILITY: | ± 0.5 °C |
TEMPERATURE HOMOGENEITY: | ± 0.5 °C |
TEMPERATURE PRECISION: | ± 1 °C |
REACHING OF SET TEMPERATURE: | acoustic and visual signal |
COUNTDOWN: | visual signal |
END OF CYCLE: | acoustic and visual signal |
PROBE INTERRUPTION: | acoustic and visual signal |
BROKEN PROBE: | acoustic and visual signal |
EXCEED TEMPERATURE RANGE: | acoustic and visual signal |
16-position thermoreactor for Chemical Oxygen Demand (COD) with settable temperature (up to 160 °C) and timer (up to 199 minutes).
COD Analysis Thermoreactor ECO 16 ensures a simple and easy programming of time (from 1 to 199 minutes or continuous) and temperature (from ambient to 160 °C).
Through the digital display, it is possible to monitor the test at any time, with the countdown constantly shown.
The unit automatically shuts down when the analysis time expires, whilst safety is of primary importance, as demonstrated by the antisplash cover that protects the user and other samples and the several acoustic and visual signals that report unpleasant situations. Eco 16 has also the transparent test tube cover that offers a constant visual control.
CONSTRUCTION MATERIAL: | Epoxy painted metal structure |
HOLES NUMBER AND DIAMETER: | 14 pos. (Ø 16 mm) and 2 pos. (Ø 22 mm) |
SET TEMPERATURE: | display visualization |
COUNTDOWN: | display visualization |
END OF THE CYCLE: | acoustic signal with automatic switch-off |
POWER: | 550 W |
WEIGHT: | 3,8 Kg (8.4 Ib) |
DIMENSIONS (WXHXD): | 168x110x269 mm (6.4×4.3×10.6 in) |
SELECTABLE WORKING TEMPERATURES: | from room temperature to 160 °C, resolution 1 °C |
SELECTABLE WORKING TIMES: | from 0 to 199 minutes or continuous operation |
TEMPERATURE STABILITY: | ± 0.5 °C |
TEMPERATURE HOMOGENEITY: | ± 0.5 °C |
TEMPERATURE PRECISION: | ± 1 °C |
REACHING OF SET TEMPERATURE: | visual signal |
COUNTDOWN: | visual signal |
END OF CYCLE | acoustic and visual signal |
PROBE INTERRUPTION: | acoustic and visual signal |
BROKEN PROBE: | acoustic and visual signal |
EXCEED TEMPERATURE RANGE: | acoustic and visual signal |