Product name / Cat. No: ADAMⅡ / ADAMⅡ
Camera: High-sensitivity monochrome CCD
Optical: Bright, Yellow (PE), Red (PerCP) and Green (FITC)
Weight: 19.3 kg
Size (LxWxH): 300 x 420 x 370 mm
Lens: Olympus lens (10X) / 35∮tube lens
Specimen: HPC-A, Mobilized PB, Control
Filter: PerCP: BLP01-635R (650 long pass)
Total frame / Total Volume: 136 frame / 7.854 uL
PE: Excitation 525/50, Emission 572/28-25
Reading range / Chamber Volume: 28 x 8 x 0.1 mm / 25 uL
FITC: Excitation 466/40, Emission 550/50
Stage: Automated X-Y-Z stage
ADAMⅡ-CD34 Kit
Principles of procedure
When the sample is mixed with reagent in the test tube, fluorescence-labeled antibodies bind specifically to the CD34+ and CD45+ cell
surface and dead cells are stained by nucleic acid staining dye.
Once the staining step is completed, RBC Lysis buffer is added for the lysis of red blood cells.
The lysed sample is loaded into a disposable plastic slide and the slide is placed on the precision stage of ADAMⅡ instrument.
Product name / Cat. No:ADAMⅡ-CD34 Kit / CD34K-025
Staining solution: PE, PerCP, Nucleic acid dye
1 Box: 25 tests
Loading sample volume: 25 µL / test
ADVANTAGE
Simple and easy to use
The ADAM-rWBC Kit consists of standard bead solution, r-solution, and r-slide.
For absolute counting of residual leukocytes, appropriate sample preparation is needed.
After mixing the sample with the provided r-Solution, load the sample into the r-Slide.
Then, the number of total leukocytes will be automatically calculated and displayed on the ADAM-rWBC series.
Product name / Cat. No: ADAM rWBC Kit / AD1K-050
Quantity: 50 tests / kit
Staining method: PI (Propidium lodide)
Sample loading volume: 100 µL
Contents: r-Slide: 50 slides / r-Solution: 25 ml x 1 bottle / Standard bead solution: 7 ml x 1 bottle